
Products
SHBC provides colored microspheres, fluorescent microspheres, magnetic beads, silica microspheres, chromatography packing microspheres and biological reagents for diagnostic assay development, nucleic acid extraction, protein purification and separation applications.
-
Cyto5UM-10
-
SHBC
-
0.5%
-
5µm
-
10ml,20ml,50ml ,500ml,1000ml
Flow Cytometry Absolute Counting Fluorescent Microspheres
SHBC Flow Cytometry Absolute Counting Fluorescent Microspheres are dual-fluorescent reference particles developed for absolute cell enumeration, flow cytometric immunophenotyping, immune-cell analysis and quantitative particle-counting research.
The Double-Fluo Microspheres feature two fluorescence detection channels:
Excitation at approximately 488 nm with emission centered near 530 nm
Excitation at approximately 633 nm with emission centered near 670 nm
The microspheres are supplied as an aqueous suspension with a solids content of 0.5%.
The dual-laser, dual-emission fluorescence profile allows the microspheres to be identified using both blue-laser and red-laser channels on compatible flow cytometers. When a known volume of microspheres with a lot-specific, assayed particle concentration is added to a known sample volume, the absolute concentration of cells or other particles can be calculated from the ratio of sample events to microsphere events.
This product is intended for research institutions, flow cytometry laboratories, immunophenotyping reagent developers, antibody companies and manufacturers requiring development samples, pilot quantities or bulk fluorescent microsphere supply.
Product Overview
Flow cytometry commonly reports the percentage of cells within a selected population. However, many research applications also require the actual number of cells per microliter, milliliter or original sample volume.
Absolute counting microspheres provide an internal particle reference for this calculation.
A known volume of counting microspheres is added directly to the cell sample. The flow cytometer then records the target-cell events and microsphere events during the same acquisition.
Because the microsphere concentration is known, the number of cells in the original sample can be calculated from the ratio between the two event populations.
Double-Fluo Microspheres provide two fluorescence responses for microsphere identification:
Green fluorescence near 530 nm under approximately 488 nm excitation
Far-red fluorescence near 670 nm under approximately 633 nm excitation
The dual-fluorescent profile can help users:
Distinguish counting microspheres from unstained cells
Separate microspheres from cellular debris
Confirm microsphere identity in two fluorescence channels
Reduce dependence on scatter parameters alone
Establish a compact and reproducible microsphere gate
Support flow cytometers equipped with blue and red lasers
Develop customized absolute-counting reagents
Improve internal acquisition and process control
The microspheres may be evaluated with whole blood, lysed whole blood, peripheral blood mononuclear cells, cultured cells, isolated immune cells and other particle suspensions.
Compatibility must be verified with the intended sample matrix, antibody panel, cytometer and data-analysis workflow.
Key Features of Double-Fluo Microspheres
Dual Excitation at 488 nm and 633 nm
The Double-Fluo Microspheres are designed for excitation using two common flow cytometer laser lines:
Approximately 488 nm blue-laser excitation
Approximately 633 nm red-laser excitation
The 488 nm excitation provides a green fluorescent response that can be detected near 530 nm.
The 633 nm excitation provides a far-red fluorescent response that can be detected near 670 nm.
This dual-laser design supports flow cytometers equipped with both blue and red laser modules and allows reagent developers to select one or both fluorescence signals for microsphere identification.
Instruments using 635 nm or 640 nm red lasers may also be evaluated. Final compatibility depends on laser power, filter configuration, detector sensitivity and the lot-specific fluorescence spectrum.
Dual Emission at 530 nm and 670 nm
The nominal fluorescence emission channels are:
Approximately 530 nm after 488 nm excitation
Approximately 670 nm after 633 nm excitation
The 530 nm signal is generally located in the green fluorescence detection region.
The 670 nm signal is located in the far-red fluorescence detection region.
The two fluorescence signals may be used individually or together according to the antibody panel and cytometer configuration.
A dual-fluorescence gate can provide an additional method for distinguishing counting microspheres from:
Unstained cells
Cell fragments
Platelets
Electronic noise
Autofluorescent events
Single-channel fluorescent particles
Nonspecific sample contaminants
The exact filter center wavelengths and bandwidths must be verified on the intended flow cytometer.
0.5% Solids Aqueous Suspension
The product is supplied with a solids content of 0.5%.
When expressed as weight per volume, 0.5% solids corresponds to approximately 5 mg of microspheres per milliliter.
The solids content is useful for:
Manufacturing process control
Formulation development
Dilution calculations
Batch-to-batch mass comparison
Customized concentration preparation
Filling-process control
However, solids content is not the same as the particle number concentration required for absolute counting.
Lot-Specific Bead Concentration
Reliable absolute counting requires a lot-specific, assayed particle number concentration stated as:
Beads/mL
Beads/μL
Beads per recommended test volume
The bead number concentration must be measured using a validated particle-counting method and documented on the product label or certificate of analysis.
This value is used in the absolute cell-count calculation.
Important controls include:
Particle number concentration
Concentration tolerance
Filling accuracy
Bottle-to-bottle consistency
Suspension homogeneity
Sampling reproducibility
Stability during storage
Stability after transportation
Commercial absolute-counting products similarly rely on calibrated or lot-specific microsphere concentrations rather than solids content alone.
Uniform Particle Size
Controlled particle size helps generate a compact microsphere population on forward-scatter and side-scatter plots.
A narrow particle-size distribution can improve:
Gating reproducibility
Event identification
Instrument-to-instrument transfer
Sampling consistency
Doublet discrimination
Batch-to-batch comparison
Automated data analysis
The nominal particle size and permitted size range should be stated in the final product specification.
Uniform Dual-Fluorescence Intensity
Consistent fluorescence intensity in both channels helps produce a clearly separated microsphere gate.
Recommended fluorescence quality attributes include:
Mean green fluorescence intensity
Green fluorescence coefficient of variation
Mean far-red fluorescence intensity
Far-red fluorescence coefficient of variation
Percentage of dual-positive microspheres
Ratio of green to far-red fluorescence
Fluorescence stability during storage
Instrument-to-instrument detectability
The two fluorescence signals should be sufficiently separated from the intended cell populations and antibody conjugates.
Internal Fluorescent Labeling
The fluorescent dyes are incorporated into the microsphere structure rather than relying only on weak surface adsorption.
Internal fluorescent labeling can help:
Reduce dye loss
Improve washing resistance
Maintain a stable outer surface
Protect the fluorescent material
Improve storage stability
Support long-term reagent development
Reduce lot-to-lot fluorescence variation
The finished product should still be protected from prolonged exposure to strong light.
Single-Platform Absolute Counting
The microspheres and target cells are acquired during the same flow cytometry run.
This allows the user to calculate cell concentration without relying exclusively on a separate hematology analyzer or external particle counter.
The single-platform method can reduce variation introduced by transferring percentage results between different instruments. Official counting-bead products from Thermo Fisher and Beckman Coulter use the same ratio-based principle.
Suitable for Reagent Development and Bulk Manufacturing
SHBC Double-Fluo Microspheres are positioned for customers requiring more than a small research package.
Available project stages may include:
Initial microsphere evaluation
Flow cytometer compatibility testing
Gating-method development
Absolute-count reagent development
Antibody-panel compatibility testing
Pilot manufacturing
Stability studies
Process validation
Bulk production
Lot reservation
OEM and ODM manufacturing
Technical Specifications
Parameter | Specification |
|---|---|
Product Name | Flow Cytometry Absolute Counting Fluorescent Microspheres |
Product Type | Double-Fluo Microspheres |
Primary Application | Flow cytometry absolute counting |
Solids Content | 0.5% |
Excitation Wavelength 1 | Approximately 488 nm |
Emission Wavelength 1 | Approximately 530 nm |
Excitation Wavelength 2 | Approximately 633 nm |
Emission Wavelength 2 | Approximately 670 nm |
Fluorescence Profile | Dual fluorescent |
Detection Method | Flow cytometry |
Product Form | Aqueous microsphere suspension |
Particle Number Concentration | Lot-specific; stated on COA |
Particle Size | According to the final product specification |
Microsphere Material | According to the final product specification |
Recommended Applications | Absolute cell enumeration, immunophenotyping and immune-cell research |
Supply Format | Evaluation, pilot and bulk quantities |
Intended Use | Research use and reagent development |
The following information should be confirmed and added before formal publication:
Catalog number
Nominal particle size
Particle-size tolerance
Microsphere matrix
Lot-specific beads/mL
Particle-count concentration tolerance
Suspension buffer
Surfactant
Preservative
Package size
Storage temperature
Shelf life
Solids Content vs. Bead Number Concentration
Solids content and particle number concentration describe different product properties.
What Does 0.5% Solids Mean?
A 0.5% weight-per-volume suspension contains approximately:
5 mg of microspheres per milliliter
Solids content describes the total mass of microsphere material in a given volume.
It is useful for manufacturing, dilution and formulation control.
What Does Beads/mL Mean?
Beads/mL describes the number of individual microspheres in one milliliter.
This value depends on:
Microsphere diameter
Particle density
Particle-size distribution
Solids concentration
Aggregate level
Particle-counting method
Two microsphere products can have the same 0.5% solids content but very different beads/mL values if their particle sizes are different.
Which Value Is Used for Absolute Counting?
The absolute cell-count formula requires the calibrated particle number concentration, not only the solids content.
The COA should therefore provide a lot-specific concentration such as:
8.5 × 10⁵ beads/mL
1.0 × 10⁶ beads/mL
A specified number of beads per recommended test volume
The examples above illustrate reporting formats only and are not product specifications.
How Absolute Counting Microspheres Work
A known volume of Double-Fluo Microspheres is added to a known volume of the sample.
During acquisition, the flow cytometer records:
The number of target-cell events
The number of counting-microsphere events
Because the number of microspheres added to the tube is known, the original concentration of target cells can be calculated from the event ratio.
A typical process includes:
Prepare or stain the cell sample.
Lyse red blood cells if required.
Mix the counting microsphere suspension thoroughly.
Add an accurately measured volume of microspheres.
Acquire the sample without losing cells or microspheres.
Gate the target-cell population.
Gate the Double-Fluo Microsphere population.
Record cell and microsphere event numbers.
Calculate the absolute cell concentration.
Apply any required sample-dilution factor.
Absolute Cell Count Calculation Formula
The general calculation is:
Absolute cell concentration = Cell events ÷ Bead events × Bead concentration × Bead volume ÷ Sample volume × Dilution factor
When the result is expressed as cells per microliter:
Cells/μL = (Cell events / Bead events) × Bead concentration in beads/μL × (Bead volume / Sample volume) × Dilution factor
Where:
Cell events are events within the validated target-cell gate.
Bead events are events within the validated Double-Fluo Microsphere gate.
Bead concentration is the lot-specific value stated on the COA.
Bead volume is the actual volume of bead suspension added.
Sample volume is the original sample volume used for the calculation.
Dilution factor accounts for any applicable dilution.
Simplified Calculation
When the bead and sample volumes are equal:
Cells/μL = Cell events ÷ Bead events × Bead concentration in beads/μL
This simplified formula is only valid when the sample-preparation design supports the assumption.
Thermo Fisher and Beckman Coulter counting-bead methods use the same underlying ratio between target-cell events and assayed microsphere events.
Why Use Double-Fluo Counting Microspheres?
Confirm Microsphere Identity in Two Channels
A counting population identified using only scatter or one fluorescent signal may overlap with sample debris or other fluorescent events.
A dual-fluorescent signature provides an additional level of microsphere identification.
Potential gating options include:
530 nm fluorescence alone
670 nm fluorescence alone
530 nm versus 670 nm
FSC versus 530 nm
SSC versus 670 nm
Dual-positive microsphere gating
Reduce Dependence on Scatter Alone
Forward and side scatter can vary with:
Cytometer model
Laser alignment
Detector voltage
Sample matrix
Particle refractive index
Threshold settings
Instrument maintenance
Fluorescence-based identification can help maintain microsphere detection when scatter separation is limited.
Support Blue- and Red-Laser Instruments
Many multicolor flow cytometers contain both blue and red lasers.
Double-Fluo Microspheres allow manufacturers to develop counting reagents compatible with:
Blue-laser detection
Red-laser detection
Dual-laser confirmation
Different filter configurations
Different antibody panels
Improve Internal Quality Control
When a defined dual-fluorescence relationship is established, users can monitor:
Loss of one fluorescence signal
Incorrect gate placement
Optical-channel failure
Sample contamination
Microsphere aggregation
Abnormal signal ratio
Instrument-setting changes
Support Automated Analysis
A compact, dual-positive microsphere population can help simplify automated identification by analysis software.
Automated gating should still be validated across:
Multiple lots
Multiple instruments
Multiple operators
Different sample matrices
Low and high cell concentrations
Flow Cytometer and Detector Compatibility
Blue-Laser Channel
The nominal blue-laser configuration is:
Excitation: approximately 488 nm
Emission: approximately 530 nm
Possible initial detector configurations include filters centered near the green fluorescence region.
The actual filter bandwidth must be selected according to the microsphere spectrum and the antibody panel.
Red-Laser Channel
The nominal red-laser configuration is:
Excitation: approximately 633 nm
Emission: approximately 670 nm
Instruments using 635 nm or 640 nm red lasers may be compatible after verification.
Antibody Panel Considerations
The 530 nm channel may overlap with common green-emitting antibody conjugates.
The 670 nm channel may overlap with far-red fluorophores.
During panel development, evaluate:
Fluorescence overlap
Spectral spillover
Compensation requirements
Detector saturation
Microsphere brightness
Cell-marker intensity
Available unused channels
The microsphere gate may be established using the channel that provides the best separation from the stained cell populations.
Conventional and Spectral Cytometers
Double-Fluo Microspheres may be evaluated on:
Conventional filter-based flow cytometers
Spectral flow cytometers
Benchtop flow cytometers
Automated flow cytometry analyzers
Cell-analysis platforms with volumetric sampling
Research cell sorters
For spectral cytometers, the full fluorescence signature should be measured and included in the reference library if appropriate.
Recommended Applications
Absolute Immune-Cell Enumeration
The microspheres may be evaluated for absolute enumeration of:
Total leukocytes
T cells
CD4-positive T cells
CD8-positive T cells
B cells
Natural killer cells
Monocytes
Granulocytes
Dendritic-cell populations
Other gated immune-cell subsets
The target population is defined using the laboratory’s validated antibody panel and gating strategy.
Flow Cytometric Immunophenotyping
Double-Fluo Microspheres can be added to immunophenotyping workflows to convert cell percentages into absolute numbers.
Potential research formats include:
Whole-blood immunophenotyping
Lyse-no-wash assays
PBMC analysis
Bone-marrow research
Cell-culture analysis
Immune-monitoring studies
Preclinical research
Cell-therapy process development
Cell Culture and Bioprocess Monitoring
Potential research applications include:
Cultured immune-cell counting
T-cell expansion monitoring
NK-cell culture monitoring
Cell-recovery studies
Cell-manufacturing process development
Cell viability and yield research
Bioprocess sampling
Viability dyes and counting microspheres can be combined when the fluorescence panel provides sufficient channel separation.
Stem and Progenitor Cell Research
The microspheres may be evaluated in research workflows involving:
CD34-positive cell enumeration
Hematopoietic progenitor research
Cell-processing recovery
Cryopreservation studies
Cell-therapy process development
Method-specific validation is required.
Microorganism and Particle Counting
After appropriate validation, the same ratio-based principle may be evaluated for:
Yeast
Bacteria
Microalgae
Synthetic particles
Extracellular vesicle preparations
Environmental particles
Particle size, scatter threshold and fluorescence separation must be suitable for the target.
Instrument and Method Development
The product may also support:
Flow cytometer method development
Sample-preparation optimization
Pipetting-method validation
Operator training
Software-gating development
Counting-precision studies
Inter-instrument comparison
OEM counting reagent development
Recommended Flow Cytometry Workflow
Step 1: Prepare the Sample
Prepare a known volume of the biological sample.
Record:
Original sample volume
Any predilution
Staining volume
Lysis volume
Final dilution factor
Any washing step
Loss of cells during washing can affect the relationship between the measured result and the original sample.
Step 2: Stain the Cells
Add the required antibody panel and incubate according to the validated method.
Protect fluorescent reagents from unnecessary light exposure.
Step 3: Lyse Red Blood Cells if Required
For whole-blood applications, use a validated lysis method.
A lyse-no-wash workflow can reduce cell loss, but compatibility with the microsphere suspension and antibody panel must be confirmed.
Step 4: Mix the Microspheres
Immediately before pipetting:
Bring the product to the required operating temperature.
Mix according to the validated instructions.
Confirm that no settled layer remains.
Avoid generating foam.
Avoid prolonged high-speed vortexing unless validated.
Inspect for visible aggregates.
AccuCheck recommends careful mixing and reverse pipetting to improve counting accuracy.
Step 5: Pipette an Accurate Microsphere Volume
Use a calibrated pipette.
For critical quantitative work, consider:
Reverse pipetting
Low-retention tips
Consistent aspiration depth
Consistent pipetting speed
Wet-tip dispensing
Gravimetric pipette verification
Replicate dispensing studies
Step 6: Acquire the Sample Promptly
After adding the microspheres:
Mix gently.
Avoid prolonged delay before acquisition.
Keep the sample protected from light.
Avoid unnecessary settling.
Maintain consistent acquisition timing.
Step 7: Collect Sufficient Microsphere Events
For development work, collecting at least 1,000 microsphere events is a useful starting target, based on the AccuCheck workflow. The final event requirement should be established from the precision needed for the specific method.
Low microsphere event numbers can increase statistical uncertainty.
Step 8: Apply the Validated Gates
Record:
Total microsphere events
Target-cell events
Excluded aggregate events
Acquisition time
Total analyzed volume, when available
Instrument settings
Step 9: Calculate the Result
Use the lot-specific microsphere concentration and the validated formula.
Do not substitute the 0.5% solids content for the particle number concentration.
Suggested Gating Strategy
Gate 1: Time Stability
Plot time against one fluorescence channel or total event rate.
Exclude periods showing:
Interrupted flow
Sample clogs
Sudden event-rate changes
Air bubbles
Microsphere settling
Sample aspiration instability
Gate 2: Broad Particle Gate
Use FSC and SSC to include:
Target cells
Counting microspheres
Exclude:
Electronic noise
Very small debris
Extreme aggregates
Do not set the threshold so high that counting microspheres are removed.
Gate 3: Microsphere Fluorescence Gate
Plot 530 nm fluorescence against 670 nm fluorescence.
Define the validated Double-Fluo Microsphere region.
The microsphere population should show a consistent relationship between the two fluorescence signals.
Gate 4: Singlet Microspheres
Where pulse geometry is available, use:
Area versus height
Area versus width
Exclude microsphere aggregates and coincident events.
Gate 5: Target-Cell Population
Apply the laboratory’s validated immunophenotyping gates.
Potential markers may include:
CD45
CD3
CD4
CD8
CD19
CD16
CD56
CD14
Viability markers
Gate 6: Final Event Review
Confirm:
Microsphere gate separation
Adequate bead event count
Stable bead fluorescence ratio
Expected cell-population distribution
Absence of major acquisition instability
Valid controls
Factors Affecting Absolute Counting Accuracy
Microsphere Concentration Accuracy
The beads/mL value directly affects the calculated cell result.
An error in the assigned microsphere concentration produces a proportional error in the calculated cell concentration.
Pipetting Accuracy
The calculation assumes that the stated volume of microspheres was actually added.
Potential sources of error include:
Pipette calibration
Tip retention
Incomplete dispensing
Air bubbles
Incorrect aspiration depth
Microsphere settling
Operator technique
Thermo Fisher notes that counting accuracy depends strongly on sample handling and precise bead-volume delivery.
Suspension Homogeneity
Microspheres may settle over time.
If the suspension is not homogeneous, different aliquots may contain different numbers of microspheres.
Aggregate Formation
Aggregates may be counted as:
One event
An excluded event
A cell-like event
An abnormal fluorescent event
Aggregate control is therefore important for accurate counting.
Event Statistics
The precision of a ratio-based calculation depends on the number of recorded cell and microsphere events.
Very low event counts increase statistical variation.
Gating Consistency
Changing the microsphere or cell gate between samples can change the calculated result.
Automated or locked analysis templates may improve consistency after validation.
Threshold Settings
A high FSC or SSC threshold can remove small counting microspheres.
A fluorescence-based threshold may be considered when compatible with the full antibody panel.
Sample Loss
Centrifugation, aspiration and washing can remove cells or microspheres.
The calculated result must be interpreted according to the sample-preparation workflow.
Coincidence and High Event Rate
Excessive event rates can cause coincident passage of multiple particles through the detection point.
Dilute the sample or reduce flow rate when required.
Instrument Stability
Changes in:
Laser power
Fluidics
Optical alignment
Detector voltage
Filter performance
Flow rate
can affect microsphere identification and counting reproducibility.
Troubleshooting Guide
Microspheres Are Not Visible
Possible causes:
Threshold is too high
Incorrect fluorescence channel
Incorrect laser selected
Detector voltage is too low
Microspheres have settled
Sample was not mixed
Fluorescence filter is incompatible
Microsphere concentration is too low
The product was exposed to damaging conditions
Recommended actions:
Reduce the scatter threshold.
Check the 530 nm and 670 nm channels.
Confirm laser availability.
Mix the suspension thoroughly.
Run microspheres without cells.
Compare with the lot-specific reference data.
Microsphere Gate Is Too Broad
Possible causes:
Aggregation
Inadequate mixing
Excessive vortexing
Contamination
Detector saturation
Mixed singlet and doublet events
Instrument instability
Recommended actions:
Apply singlet gating.
Reduce detector gain if saturated.
Inspect the suspension.
Use clean tubes and filtered buffers.
Compare the result with a retained reference lot.
Green Signal Is Strong but Far-Red Signal Is Weak
Possible causes:
Red laser is unavailable or incorrectly configured
Far-red detector voltage is too low
Filter center wavelength is unsuitable
Red-laser alignment is poor
Fluorescence has changed during storage
Incorrect analysis channel was selected
Far-Red Signal Is Strong but Green Signal Is Weak
Possible causes:
Blue laser is unavailable or incorrectly configured
Green detector voltage is too low
Green channel is occupied or compensated incorrectly
Incorrect emission filter is installed
The 530 nm signal is outside the acquisition scale
Cell Count Is Higher Than Expected
Possible causes:
Microsphere volume was under-delivered
Bead concentration value was entered incorrectly
Microspheres were excluded from the gate
Aggregates were excluded as one event
Sample dilution factor was incorrect
Cell gate included debris or doublets
Cell Count Is Lower Than Expected
Possible causes:
Excessive microsphere volume
Incorrectly high beads/mL value
Cell loss during washing
Target cells excluded from the gate
Cell aggregates counted as single events
Incorrect original sample volume
Sample degradation
Results Vary Between Replicates
Possible causes:
Microsphere settling
Inconsistent mixing
Pipetting variation
Different acquisition delays
Low event counts
Gate movement
Variable flow rate
Sample heterogeneity
Double-Fluo Beads vs. Other Counting Methods
Counting Method | Principle | Main Advantages | Main Considerations |
|---|---|---|---|
Double-Fluo Counting Microspheres | Ratio of cell events to dual-fluorescent bead events | Dual-channel bead identification and single-platform counting | Requires calibrated beads/mL and accurate pipetting |
Broad-Spectrum Counting Beads | Ratio using broadly fluorescent beads | Compatible with many lasers and detectors | May occupy multiple channels |
Two-Population Counting Beads | Two different bead populations as internal standards | Additional check of sampling and mixture proportion | More complex gating |
Volumetric Flow Cytometry | Instrument measures analyzed sample volume | No external counting beads required | Depends on instrument volume calibration |
Hematology Analyzer Plus Flow Cytometry | Cell concentration from one instrument and percentages from another | Familiar laboratory workflow | Multiple-platform variation |
Manual Hemocytometer | Direct visual counting | Low equipment requirement | Operator-dependent and lower throughput |
Automated Cell Counter | Imaging or impedance-based counting | Fast total-cell measurement | May not identify specific immune subsets |
Double-Fluo Microspheres are particularly useful when users require an internally added counting standard with two fluorescence-identification channels.
Quality Control for Batch Production
Absolute counting microspheres require both material quality control and metrological control.
Recommended release parameters include:
Physical Properties
Mean particle size
Particle-size distribution
Coefficient of variation
Particle morphology
Aggregate percentage
Density, when relevant
Appearance
Redispersion performance
Fluorescence Properties
488 nm excitation response
530 nm emission intensity
633 nm excitation response
670 nm emission intensity
Dual-positive percentage
Green fluorescence CV
Far-red fluorescence CV
Green-to-far-red intensity ratio
Photostability
Storage stability
Concentration Properties
Solids content
Assayed beads/mL
Particle-count tolerance
Bottle-to-bottle variation
Filling-volume accuracy
Sampling homogeneity
Stability of assigned concentration
Functional Properties
Microsphere gate separation
Recovery during flow cytometry
Agreement with reference-count samples
Linearity across cell concentrations
Within-run precision
Between-run precision
Operator reproducibility
Instrument compatibility
Recommended Incoming Inspection
Reagent manufacturers should consider testing each incoming lot for:
Appearance
Redispersion
Particle size
Aggregate level
Green fluorescence
Far-red fluorescence
Dual-positive percentage
Beads/mL
Flow cytometry recovery
Comparison with a qualified reference lot
Reference-Lot Strategy
A qualified reference lot should be retained for comparison with future lots.
New lots can be evaluated using:
The same cytometer
The same detector settings
The same sample
The same pipette
The same acquisition template
The same gating method
The same calculation software
Batch Documentation
Available or customized documents may include:
Certificate of Analysis
Product specification
Safety Data Sheet
Particle-size report
Fluorescence report
Bead-concentration report
Stability data
Lot-comparison data
Manufacturing change notification
Customization and Bulk Manufacturing
SHBC supports customized fluorescent microsphere development for flow cytometry reagent manufacturers.
Potential customization options include:
Particle size
Particle-size distribution
488 nm excitation intensity
530 nm emission intensity
633 nm excitation intensity
670 nm emission intensity
Green-to-far-red intensity ratio
Solids content
Beads/mL
Suspension buffer
Surfactant system
Preservative system
Package volume
Low-binding formulation
OEM labeling
Private-label packaging
Development Supply
Small quantities can be supplied for:
Instrument compatibility
Gating development
Antibody-panel evaluation
Pipetting studies
Concentration optimization
Stability screening
Pilot Production
Pilot lots can support:
Process transfer
Filling validation
Method validation
Inter-instrument comparison
Packaging evaluation
Accelerated stability studies
Bulk Manufacturing
Bulk production may include:
Lot reservation
Customer-specific release criteria
Customized concentration
Customized packaging
Long-term supply planning
OEM and ODM services
Batch documentation
Storage and Handling
Recommended practices include:
Store at 2–8°C unless otherwise stated on the product label or COA.
Protect the product from direct light.
Do not freeze unless freeze-thaw stability has been validated.
Mix thoroughly before use.
Avoid contamination of the stock suspension.
Use calibrated pipettes.
Use clean, low-binding tubes and tips.
Avoid prolonged exposure to elevated temperature.
Close the bottle immediately after use.
Follow the lot-specific expiration date.
Commercial counting-bead products are commonly stored at 2–8°C and protected from light.
Before use:
Inspect the suspension.
Confirm that settled microspheres can be redispersed.
Mix according to the validated procedure.
Avoid introducing foam.
Pipette immediately after mixing.
Return the bottle to the recommended storage condition.
Frequently Asked Questions
What are Double-Fluo Absolute Counting Microspheres?
They are fluorescent reference microspheres designed for absolute cell enumeration by flow cytometry.
The microspheres provide fluorescence responses near 530 nm and 670 nm under approximately 488 nm and 633 nm excitation.
What is the solids content?
The product has a solids content of 0.5%.
This corresponds to approximately 5 mg of microsphere material per milliliter when expressed as weight per volume.
Can 0.5% solids be used directly in the counting formula?
No.
The absolute-counting formula requires the lot-specific particle number concentration in beads/mL or beads/μL.
Solids content alone is insufficient.
Why does the product need a lot-specific beads/mL value?
The calculated cell concentration is directly proportional to the assigned microsphere concentration.
Any error in beads/mL produces a corresponding error in the cell-count result.
What lasers are required?
The nominal excitation wavelengths are 488 nm and 633 nm.
Flow cytometers using a 635 nm or 640 nm red laser may also be evaluated.
Which fluorescence channels are used?
The nominal emission channels are approximately 530 nm and 670 nm.
The exact detector filters must be matched to the actual fluorescence spectra.
Can the microspheres be detected using only one channel?
Potentially yes.
However, using both fluorescence channels may improve microsphere identification and gating confidence.
The method should be validated with the intended antibody panel.
Can Double-Fluo Microspheres be used with FITC- or APC-conjugated antibodies?
They may overlap with green- or far-red-emitting antibody conjugates.
Panel compatibility, compensation and gate separation must be evaluated before use.
Can the microspheres be used with whole blood?
They may be evaluated with whole blood and lyse-no-wash workflows.
The lysis reagent, antibody panel, sample volume and cytometer must be validated.
Can they be used with cultured cells?
Yes. They may be evaluated for absolute enumeration of cultured cells, immune cells and other particle suspensions.
How many bead events should be collected?
A minimum of approximately 1,000 bead events can be used as a starting target during method development.
The final requirement should be based on the required precision and expected sample concentration.
Why can I not see the microspheres on FSC and SSC?
The scatter threshold may be too high, or the microspheres may have lower scatter than the target cells.
Reduce the threshold carefully and use the fluorescence channels to identify the microsphere population.
Why do the microspheres settle?
Microspheres have a different density from the suspension medium and may gradually settle during storage.
Mix thoroughly before every aliquot.
Can the particle size be customized?
Particle-size customization may be available according to project quantity and technical requirements.
Can the fluorescence intensity be customized?
Green intensity, far-red intensity and the ratio between the two channels may be discussed for OEM projects.
Can the solids content be customized?
Yes. Different solids contents may be available depending on production quantity and project requirements.
Is bulk supply available?
Yes. SHBC supports evaluation quantities, pilot batches, bulk manufacturing, lot reservation and customized packaging.
Is this a finished diagnostic reagent?
No.
The microspheres are supplied for research use and reagent development. They are not a finished diagnostic product unless separately validated and registered by the customer.
Request a Sample or Bulk Quotation
SHBC supplies Double-Fluo Flow Cytometry Absolute Counting Fluorescent Microspheres for immune-cell enumeration, immunophenotyping research, flow cytometer method development and batch reagent manufacturing.
To request a sample, technical consultation or bulk quotation, please provide:
Intended application
Target cell population
Sample type
Flow cytometer model
Available laser wavelengths
Detector filter configuration
Antibody panel
Required particle size
Required beads/mL
Required solids content
Required package size
Expected annual demand
Special quality-control requirements
Contact SHBC to evaluate Double-Fluo Microspheres for your absolute cell counting, immune-monitoring, flow cytometry or OEM reagent-development project.


